Some EpiPen and EpiPen Jr auto-injectors may be difficult to remove from their carrier tubes

**** Health Canada Release

Some EpiPen and EpiPen Jr auto-injectors may be difficult to remove from their carrier tubes

20191015_155250

Pfizer Canada has advised Health Canada that, in a very small number of cases, some EpiPen (0.3 mg) and EpiPen Jr (0.15 mg) auto-injector devices may not slide out of their carrier tube easily, or at all. Although the chance of this occurring is very rare, failure to administer epinephrine as soon as possible during an anaphylactic response could lead to patient disability or death. The number of EpiPen or EpiPen Jr auto-injectors with this issue is estimated to be two auto-injectors out of every one million.

According to the company, a very small number of carrier tubes have a defect that partially blocks the opening of the tube and may prevent easy release of the auto-injector from the carrier. The issue is with the carrier tube, and not with the device itself or the drug that it delivers (epinephrine). The company has identified the root cause of this issue and put corrective measures in place.

Products are not being recalled by Pfizer because the risk can be easily addressed by consumers and pharmacists checking devices—before an emergency situation arises—to make sure they slide easily out of their carrier tube. Should you have a device with this defect, please return it to your pharmacist for replacement.

EpiPen and EpiPen Jr are used to deliver an emergency treatment of adrenaline (epinephrine) to patients who are at risk or have a history of life-threatening allergic reactions (anaphylaxis).

According to the company, there have been no reports of malfunctions related to this specific issue in Canada.

Affected products

EpiPen (0.3 mg) (DIN 00509558) and EpiPen Jr (0.15 mg) (DIN 00578657) products expiring on or before September 2020

What consumers should do

Pfizer Canada is advising consumers to check whether their device expires on or before September 2020, and if so, to do the following (images are provided below):

  • Check your device to make sure that it slides out easily from the carrier tube. To check the device, flip open the carrier tube cap, gently turn the tube upside down and let the device slide out into your hand (do not shake or drop it). Visually inspect the carrier tube to make sure that the open rim of the carrier tube is not deformed.
    • Do NOT remove the blue safety release from the auto-injector device. The blue safety release should be kept on the auto-injector until the time of use.
    • Return the EpiPen device to its carrier tube and close the carrier tube flip cap once you have confirmed that your device slides out easily and that there are no deformities on the open rim of the carrier tube.
  • If your device does not slide out easily from its tube, or the open rim of the plastic carrier tube appears deformed, return it to your pharmacist for replacement.
  • If you are unsure about how to check your device, your pharmacist can check your device for you.

In addition, Health Canada recommends that you:

  • Develop a habit of checking a new auto-injector when it is received to ensure the device can be easily removed from its packaging. Although problems with removal are rare, they have occurred on previous occasions and may reoccur periodically.
  • Report adverse events involving health products to Health Canada by calling toll-free at 1‑866‑234‑2345, or by reporting online, by mail or by fax.
  • Report complaints about health products to Health Canada by calling toll-free at 1‑800‑267‑9675, or complete an online complaint form.

Additional information for pharmacists and other health professionals:
Additional information for pharmacists and other healthcare professionals has been communicated by Pfizer Canada.